PSG vs. ROC Curve
Power Source: | Electricity | Warranty: | 1 Year |
---|---|---|---|
After-sale Service: | Online Technical Support | Power Supply Mode: | Built-in Battery |
Material: | Plastic | Shelf Life: | 1years |
Quality Certification: | Ce | Instrument Classification: | Class II |
Safety Standard: | None | Color: | White +Blue |
User: | Adult And Child | Application: | Home Sleep Test For Sleep Apnea |
Certificate: | CE/ ISO13485 | Usage: | All Kinds Of People |
Weight: | 50g (exclusive SpO2 Sensor) | APP: | Android :Search "SleepCare Plus" |
Deliver Y Time: | Samples: About 2 Days | Spo2: | 35%-100% |
Product Name: | Pulse Oximeter | ||
Highlight: | Sleep Aid Wrist Pulse Oximeter,Plastic Wrist Pulse Oximeter,Wearable Wrist Sleep Aid Device |
Display SpO2, PR
|
Battery Type: 3.7v (1000mAh) Rechargeable Lithium Batteries
|
Pulse Rate:
Measurement range: 25 - 250bpm Accuracy: ±2bpm |
SpO2:
Measurement range: 35% - 100% Accuracy: ±2% (80% - 100%) ;±3% (70% - 79%) |
Working Voltage: DC 3.4v ~ DC 4.3v
|
Free App: “BerrySmartHealth”
|
Size:68(L)x58(W)x20(D)mm
|
Weight: 50g
|
Comparison: Reference Pulse Oximeter
|
|
|
|
|
|
|||||
Berry Pulse Oximeter
|
70 ~100
|
90 ~ 100
|
80 - < 90
|
67 - < 80
|
|
|||||
# pts
|
240
|
79
|
81
|
81
|
|
|||||
Error
|
-0.6
|
-1.1
|
-0.9
|
0.1
|
Pass
|
|||||
Root mean square error
|
1.9
|
1.7
|
2.0
|
2.1
|
|
Mean Blood Oxygen
|
|
|
||
Average blood oxygen calculation accuracy: 98.51%
|
|
|
||
Diagnosis of hypoxemia
|
|
|
||
Sensitivity: 89.29%
|
Specificity: 77.78%
|
Accuracy:92.59%
|
||
OSAHS Judgment Results
|
|
|
||
|
|
|
||
Sensitivity: 94.17%
|
Specificity: 61.21%
|
Accuracy:89.62%
|
Display SpO2, PR
|
Battery Type: 3.7v (1000mAh) Rechargeable Lithium Batteries
|
Pulse Rate:
Measurement range: 25 - 250bpm Accuracy: ±2bpm |
SpO2:
Measurement range: 35% - 100% Accuracy: ±2% (80% - 100%) ;±3% (70% - 79%) |
Working Voltage: DC 3.4v ~ DC 4.3v
|
Free App: “BerrySmartHealth”
|
Size:68(L)x58(W)x20(D)mm
|
Weight: 50g
|
Comparison: Reference Pulse Oximeter
|
|
|
|
|
|
|||||
Berry Pulse Oximeter
|
70 ~100
|
90 ~ 100
|
80 - < 90
|
67 - < 80
|
|
|||||
# pts
|
240
|
79
|
81
|
81
|
|
|||||
Error
|
-0.6
|
-1.1
|
-0.9
|
0.1
|
Pass
|
|||||
Root mean square error
|
1.9
|
1.7
|
2.0
|
2.1
|
|
Mean Blood Oxygen
|
|
|
||
Average blood oxygen calculation accuracy: 98.51%
|
|
|
||
Diagnosis of hypoxemia
|
|
|
||
Sensitivity: 89.29%
|
Specificity: 77.78%
|
Accuracy:92.59%
|
||
OSAHS Judgment Results
|
|
|
||
|
|
|
||
Sensitivity: 94.17%
|
Specificity: 61.21%
|
Accuracy:89.62%
|